NaCl% Percent Scale
The NaCl % scale is an older scale salinity scale used for seawater salinity. In
this scale 100% salinity is equivalent to roughly 10% solids. High percentages
were made by evaporation. To display NaCl in % units, enter SETUP and select
NaCl% unit. Press the RANGE key until “NaCl%” is displayed on the LCD.
The instrument will display the NaCl% reading on the 1st LCD line and the
temperature reading on the 2nd LCD line.
If the reading is out of range, the full-scale value
(400.0%) will be displayed blinking.
f(t)=
T-15
1 + 0.0162•(t-15)
a
0
= 0.008
a
1
= -0.1692
a
2
= 25.3851
a
3
= 14.0941
a
4
= -7.0261
a
5
= 2.7081
c
0
= 0.008
c
1
= 0.0005
X = 400R
T
Y = 100R
T
b
0
= 0.0005
b
1
= -0.0056
b
2
= -0.0066
b
3
= -0.0375
b
4
= 0.0636
b
5
= -0.01442
R
T
- ratio of sample conductivity to standard conductivity at Temp =(T)
CT(sample)-conductivity at T ºC;
C(35,15)=42.914
µ
S/cm - the corresponding of KCI solution containing a mass
of 32.4356 g KCl/1 Kg solution
r
T
- Temperature compensation polynomial
Salinity
Measurements