Operational Guide
72
Calibration Messages
If the reading is outside limits, “WRONG STANDARD“
message will be displayed.
If the temperature goes out of (0.0 - 50.0 ºC)
range during calibration, the “WRONG STANDARD
TEMPERATURE” message will be displayed and
temperature value will blink.
GLP refers to a quality control function used to ensure uniformity of probe
calibrations and measurements. The dedicated GLP key opens a file of
the latest calibration information. Use the
q p
keys to scroll the stored
information. This includes the standards used, temperature of the
standard, Altitude and Salinity factor, time and date of the last calibration, the
expired calibration information and the probe serial number. This information
is also included with every logged data.
Last DO Calibration Data
The last DO calibration data is stored automatically
after a successful calibration. To view the DO
calibration data, press GLP when the instrument is
in measurement mode.
The instrument will display:
The calibration standard and the calibration
temperature:
0.0 % if instrument was calibrated in this point.
100% calibration point, if instrument was calibrated
in water saturated .
DO GLP
Information
DO
Calibration
Messages