11
To further reduce the effect of glass imperfections, the cuvette can be indexed and use this new index as the position mark.
For indexing one cuvette or matching multiple cuvettes, the continuous reading mode is suggested.
In this mode multiple successive readings are taken without turning off the lamp. The turbidity is
immediately displayed, reducing considerably the measurement time.
Note
: The instrument can not perform continuous readings if the average mode is on.
In order to index a cuvette follow the next steps:
• Fill the cuvette with high quality water (<0.1
FNU) up to the 10 mL mark.
• Clean and oil the cuvette as described in the
previous section.
• Turn the instrument ON.
• Insert the cuvette into the instrument and press
“
Read
” functional key. Record the reading.
• Open the instrument lid, slightly rotate the cu-
vette and take a new reading.
• Repeat the previous step until you read the lowest value.
• Alternatively, press and hold the “
Read
” key to make continuous readings. After the first value
is displayed, open the lid and start rotating the cuvette until the lowest value is displayed.