12
• Mark this position on the white band at the top of
the cuvette with a water resistant pencil.
• Always use this position to align the cuvette
with the mark on the cuvette holder.
In order to match multiple cuvettes follow the next
steps:
• Fill multiple cuvettes with high quality water
(<0.1FNU) up to the 10 mL mark.
• Clean and oil the cuvettes as described before.
• Turn the instrument ON.
•
Insert the first cuvette into the instrument and
press “
Read
” functional key. Record the
reading.
•
Open the instrument lid, slightly rotate the
cuvette and take a new reading.
•
Repeat the previous step until you read the
lowest value.
•
Mark this position on the white band at the top
of the cuvette with a water resistant pencil.
MATCHING MULTIPLE CUVETTES
Precise measurements require the use of a single cuvette. If it is not possible, cuvette selection
and matching must be performed before taking measurements.