24
FOUR-POINT CALIBRATION
•
Remove the third standard cuvette.
•
Place the cuvette for third calibration point:
•
750 FNU (NTU) calibration cuvette for NTU
ratio mode.
•
750 NTU formazin prepared standard for NTU
non-ratio and FNU modes.
•
2000 NTU formazin prepared standard for FAU
mode.
Note:
If necessary, press
UP
or
DOWN
key to edit each calibration point value to match the exact value
of the standard as measured with a reference turbidimeter.
•
Close the lid and press
“Read”
functional key. The display will show the value blinking and the
light icon during
measurement.
•
The “Store...” message is displayed for FAU, FNU and NTU non-ratio modes and the calibration
is saved. The instrument returns to the main screen.
•
The fifth proposed calibration point, 2000 NTU for NTU Ratio mode, is displayed.
•
At this moment it is possible to exit calibration by pressing
“End”
functional key.
•
If the calibration is terminated, the display will
briefly show
“
Store...
”
and the four point calibration
is saved. The instrument returns in the main screen.
FIFTH CALIBRATION POINT (NTU Ratio Turbidity only)
•
Remove the fourth standard cuvette.
•
Place the 2000 NTU calibration cuvette (or the
fifth prepared formazin standard) into the holder,
with the cuvette mark aligned to the sign on the
instrument top.