11
10
HI 93531R
HI 93532R
Keyboard Functions:
ON/OFF
: turn the meter ON and OFF
T1/T2
(
HI 93532R
only) : select reading channel (T1, T2 or T1-T2)
HOLD
: freeze the reading on display
CLR
: clear the HI and LO values
DISP
: select Normal, Relative or Average mode.
ALT
: enable the second function keys; the "ALT" tag turns on to
indicate that the second functions are enabled
Note
: The
ALT
key can be released before pressing the second
function key for 1-hand operation.
ALT/ CAL
: press and hold for about 5 seconds to enter the
Calibration mode (with reading within ±3°C range).
(ALT +) AOFF
: set auto power off delay (8 or 60 min, or disabled).
(ALT +) °C/°F
: change reading units (°C or °F).
(ALT +) MR
: recall memorized value.
(ALT +) BL
: toggle the backlight ON and OFF.
SPECIFICATIONS OF
HI 93531R & HI 93532R
Main features:
• Two independent measurement channels (
HI 93532R
only)
• Use
HI 766
K-thermocouple probes
• Display of temperature in degrees Celsius or Fahrenheit
• Extended range -200 to 1371°C or -328 to 2500°F
• Relative and average measurements
• HOLD function
• HI & LO (Max & Min) temperature values displayed on the LCD
• Remaining battery life indication / low battery detection
• Waterproof casing
• Auto-off capability
• Backlight feature
• User calibration at 0°C
• RS232 serial port communication
SPECIFICATIONS
Range (*)
-200.0 to 999.9°C / 1000 to 1371°C
-328.0 to 999.9°F / 1000 to 2500°F
Resolution
0.1°C (-149.9 to 999.9°C)
0.2°C (-200.0 to -150.0°C) / 1°C (outside)
0.1°F (-24.9 to 999.9°F)
0.2°F (-249.9 to -25.0°F)
0.3°F (-328.0 to -250.0°F) / 1°F (outside)
Accuracy
±0.5°C (-100.0 to 999.9°C) / ±1°C (outside)
(@20°C/68°F)
±1°F (-148.0 to 999.9°F) / ±1.5°F (outside)
excluding probe error
Typical EMC
±3°C / ±6°F
Deviation
with
HI 766
K-thermocouple probe
Battery
3x1.5V AA (IEC LR6) batteries,
approx. 500 hours of continuous use (BL off)
Auto-off
user selectable: 8 min, 60 min or disabled
Serial Port
unidirectional, 8-bit data, 1200 baud
Communication
(see "Serial Communication" section)
Environment
-10 to 60°C (14 to 140°F); RH 100%
Dimensions
150 x 80 x 36 mm (5.9 x 3.1 x 1.4")
Weight
235 g (8.3 oz.)
(*)
Range may be limited by probe.