background image

•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display 
shows “

Add

”, “

C.1

” with “

Press

” blinking, 

the meter is ready.

•  Fill the cuvette with 10 mL of unreacted 

sample and replace the cap. Place the cuvette 
into the meter and close the meter’s cap.

•  Press the button. When the display shows 

Add

”, “

C.2

” with “

Press

” blinking the 

meter is zeroed.

•  Remove the cuvette from the meter and 

unscrew the cap. Add the content of one 
packet of 

HI736-0

 reagent. Replace the cap 

and shake gently for 2 minutes until the 
powder is completely dissolved. Place the 
cuvette back into the meter.

•  Press and hold the button until the timer is 

displayed on the LCD (the display will show 
the countdown prior to the measurement) or, 
alternatively, wait for 3 minutes and press 
the button.

•  The instrument directly displays the 

concentration of phosphorus in ppb. The 
meter automatically turns off after 2 minutes.

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you 
use the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Measurement Procedure

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Functional Description

Errors and Warnings

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at 
tech@hannainst.com. 
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI736

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0 to 200 ppb

Resolution

1 ppb

Accuracy

±5 ppb ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the Standard Methods for the Examination of Water 
and Wastewater, 20

Th

 edition, Ascorbic Acid method. The reaction 

between phosphorus and the reagent causes a blue tint in the 
sample.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 3 minutes of non-use or 2 minutes after reading

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications