19
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:
Do not validate the instrument with any standard solutions other than the Hanna Instruments
CAL Check
TM
Standards, otherwise erroneous results will be obtained.
Note: The validation is performed only for the selected parameter. For full validation of the
instrument, the following procedure must be performed for each parameter.
• Turn the meter on by pressing
ON/OFF
.
• When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
• Place the
CAL Check
TM
Standard Cuvette A into the cuvette holder and ensure that the notch
on the cap is positioned securely into the groove.
VALID
ATION PROCEDURE
• Press
ZERO/CFM
and the lamp, cuvette and detector icons will appear on the display, depending
on the measurement phase.
• After a few seconds, the display will show
“-0.0-”
. The meter is now zeroed and ready for
validation.