background image

INSTRUCTION MANUAL

HI96738

Chlorine Dioxide ISM

13•

 When the correct year has been set, press 

ZERO/CFM

 or 

READ

 to confirm. Now 

the display will show the month blinking.

14•

 Press 

GLP/

 to edit the desired month 

(01-12). If the key is kept pressed, the 
month number is automatically increased.

15•

 When the correct month has been set, press 

ZERO/CFM

 or 

READ

 to confirm. Now 

the display will show the day blinking.

16•

 Press 

GLP/

 to edit the desired day   

(01-31). If the key is kept pressed, the 
day number is automatically increased.

Note: It is possible to change the editing 

from day to year and to month by pressing 

READ .

17•

 Press 

ZERO/CFM

 

to save the calibration 

date.

18•

 The instrument displays “

Stor

” for one 

second and the calibration is saved.

19•

 The instrument will return automatically to 

measurement mode by displaying dashes 
on the LCD.

B

HI96738

3

4-5

7

9

3

4

5-6

8

B

HI96738

11-13

14-15

16

17

Last Calibration Date

1•

 Press 

GLP/

 to enter GLP mode. The 

calibration month and day will appear 

on the main display and the year on the 

secondary display.

2•

  If no calibration was performed, the factory 

calibration message, 

“F.CAL”

 will appear on 

the main display and the instrument returns 

to measurement mode after three seconds. 

In GLP mode, the last calibration date can be verified and the factory 

calibration can be restored.

3-4

2

3-4

To save the battery, the instrument shuts down after 10 minutes of non-use 
in measurement mode and after 1 hour of non-use in calibration mode. 
If a valid measurement was displayed before 
auto-shut off, the value is displayed when the 
instrument is switched on. The blinking 

“ZERO” 

means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the 
light level.
The remaining battery capacity is evaluated at the instrument startup and 
after each measurement.
The instrument displays a battery indicator with three levels as follows:
•  3 lines for 100 % capacity
•  2 lines for 66 % capacity
•  1 line for 33 % capacity
•  Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any 
more, the instrument shows 

“dEAd bAtt”

 and turns off.

To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
•  Turn the instrument off by pressing 

ON/OFF

.

•  Turn the instrument upside down and remove the battery cover by 

turning it counterclockwise.

•  Extract the battery from its location and replace it with a fresh one.
•  Insert back the battery cover and turn it clockwise to close.

Calibration

Note: It is possible to interrupt the calibra-

tion procedure at any time by pressing  

CAL CHECK or ON/OFF keys.

 

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready.

3•

  Press and hold 

CAL CHECK 

for three 

seconds to enter calibration mode. The 

display will show 

“CAL”

 during calibration 

procedure. The blinking 

“ZERO”

 asks for 

instrument zeroing.

4•

  Place the 

CAL Check™

 Standard 

HI96738-11

 Cuvette A into the cuvette 

holder and ensure that the notch on the 

cap is positioned securely into the groove.

5•

 Press 

ZERO/CFM

 and the lamp, cuvette and 

detector icons will appear on the display, 

depending on the measurement phase.

6•

  After a few seconds the display will show     

“-0.0-”

. The meter is now zeroed and ready 

for calibration. The blinking 

“READ”

 asks 

for reading calibration standard.

7•

  Remove the cuvette.

8•

  Place the 

CAL Check™

  Standard                

HI96738-11

 Cuvette B into the holder 

and ensure that the notch on the cap is 

positioned securely into the groove.

9•

 Press 

READ

 and the lamp, cuvette and 

detector icons will appear on the display, 

depending on the measurement phase.

10•

 The instrument will show for three seconds 

the 

CAL Check™

 standard value.

Note: If the display shows “STD HIGH”, the 

standard value was too high. If the display 

shows “STD LOW”, the standard value 

was too low. Verify that both CAL Check™ 

Standard HI96738-11 Cuvettes, A and B 

are free of fingerprints or dirt and that they 

are inserted correctly.

11•

 Then the date of last calibration (e.g.: 

“01.08.2009”) appears on the display, 
or “01.01.2009” if the factory calibration 
was selected before. In both cases the year 
number is blinking, ready for date input.

12•

 Press 

GLP/

 to edit the desired year 

(2009-2099). If the key is kept pressed, 
the year number is automatically increased. 

3•

 Press 

ZERO/CFM

 to restore the factory 

calibration or press 

GLP/

 again to abort 

factory calibration restore. 

4•

  The instrument briefly indicates 

“donE” 

upon restoration of factory calibration prior 

to returning to measurement mode.

2

1

18

8

9-10

Validation

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready.

3•

  Place the 

CAL Check™

 Standard 

HI96738-11

 Cuvette A into the holder 

and ensure that the notch on the cap is 
positioned securely into the groove.

4•

 Press 

ZERO/CFM

 and the lamp, cuvette and 

detector icons will appear on the display, 
depending on the measurement phase.

5•

  After a few seconds the display will show      

“-0.0-”

. The meter is now zeroed and ready 

for validation.

6•

  Remove the cuvette.

7•

  Place the 

CAL Check™

  Standard                

HI96738-11

 Cuvette B into the holder 

and ensure that the notch on the cap is 
positioned securely into the groove.

8•

 Press 

CAL CHECK

 key and the lamp, 

cuvette and detector icons together with 

“CAL CHECK”

 will appear on the display, 

depending on the measurement phase.

9•

  At the end of the measurement the display 

will show the validation standard value.  
The reading should be within specifica-
tions as  reported on the 

CAL Check™

 

Standard Certificate. If the value is found 
out of specifications, please check that the 
cuvettes are free of fingerprints, oil or dirt 
and repeat validation. If results are still 
found out of specifications then recalibrate 
the instrument.

Warning:

 do not validate or calibrate the instrument with standard solu-

tions other than the Hanna Instruments 

CAL Check™

 Standards, otherwise 

erroneous results will be obtained.

For accurate validation and calibration results, please perform tests at room 

temperature (18 to 25 °C; 64.5 to 77.0 °F).

VALIDATION AND CALIBRATION PROCEDURES

 

Use the Hanna Instruments CAL Check™ cuvettes (see 
“Accessories”) to validate or calibrate instruments.

Validation

 

Calibration

 

GLP

Last Calibration 
Date

 

BATTERY MANAGEMENT

 or

 or

Factory Calibration 
Restore

3''

Before using these products, make sure that they are entirely suitable for your 
specific application and for the environment in which they are used.
Operation of these instruments may cause  unacceptable interferences to other 
electronic equipments, this requiring the operator to take all necessary steps 
to correct interferences. 
Any variation introduced by the user to the supplied equipment may degrade 
the instrument’s EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. 
For yours and the instrument safety do not use or store the instrument in  
hazardous environments. 

RECOMMENDATIONS FOR USERS

Factory Calibration Restore

It is possible to delete the calibration and 
restore factory calibration.

1•

 Press 

GLP/

 to enter GLP mode.

2•

 Press 

READ

 to enter in the factory cali-

bration restore screen. The instrument asks 

for confirmation of user calibration delete.

Summary of Contents for HI96738

Page 1: ...INSTRUCTION MANUAL HI96738 Chlorine Dioxide ISM ...

Page 2: ... technical support contact your local Hanna Instruments Office or e mail us at tech hannainst com Find your local Hanna Instruments Office at www hannainst com PRELIMINARY EXAMINATION For more details about spare parts and accessories see Accessories SPECIFICATIONS Range 0 00 to 2 00 mg L Resolution 0 01 mg L Accuracy 25 C 77 F 0 10 mg L 5 of reading Light source Tungsten Lamp Light Detector Silic...

Page 3: ...OFF key to turn the meter on and off 6 Liquid Crystal Display LCD 7 Cuvette alignment indicator 8 Cuvette holder 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state of the battery 4 The hourglass appears when an internal check is in progress 5 Status messages 6 The ch...

Page 4: ... mix Cylinder 2 is the sample 8 Fill a cuvette up to the 10 mL mark with blank 1 and replace the cap 9 Place the blank 1 into the holder and ensure that the notch on the cap is positioned securely into the groove 10 Press ZERO CFM and the lamp cuvette and detector icons will appear on the display depending on the measurement phase After a few seconds the display will show 0 0 The meter is now zero...

Page 5: ...sive heat greater than 25 C 78 F agitation and exeposure to light must be avoided INTERFERENCES Interference may be caused by strong oxidants 12 13 Standard Low The standard reading is less than expected Standard High The standard reading is higher than expected Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measureme...

Page 6: ...s to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase WARRANTY IST96738 01 18 Hann...

Page 7: ...ector icons together with CAL CHECK will appear on the display depending on the measurement phase 9 At the end of the measurement the display will show the validation standard value The reading should be within specifica tions as reported on the CAL Check Standard Certificate If the value is found out of specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat ...

Page 8: ... The blinking READ asks for reading calibration standard 7 Remove the cuvette 8 Place the CAL Check Standard HI96738 11 Cuvette B into the holder and ensure that the notch on the cap is positioned securely into the groove 9 Press READ and the lamp cuvette and detector icons will appear on the display depending on the measurement phase 10 The instrument will show for three seconds the CAL Check sta...

Page 9: ...is saved 19 The instrument will return automatically to measurement mode by displaying dashes on the LCD 14 15 16 17 Last Calibration Date 1 Press GLP to enter GLP mode The calibration month and day will appear on the main display and the year on the secondary display 2 If no calibration was performed the factory calibration message F CAL will appear on the main display and the instrument returns ...

Page 10: ...ps Turn the instrument off by pressing ON OFF Turn the instrument upside down and remove the battery cover by turning it counterclockwise Extract the battery from its location and replace it with a fresh one Insert back the battery cover and turn it clockwise to close 3 Press ZERO CFM to restore the factory calibration or press GLP again to abort factory calibration restore 4 The instrument briefl...

Reviews: