11
Temperature
Range
‑5.00 to 55.00 °C;
23.00 to 131.00 °F;
268.15 to 328.15 K
Resolution
0.01 °C; 0.01 °F; 0.01 K
Accuracy
± 0.15 °C; ± 0.27 °F; ±0.15 K
Calibration
Automatic at 1 custom point
pH/mV
Range
0.00 to 14.00 pH; ± 600.0 mV
Resolution
0.01 pH; 0.1 mV
Accuracy
± 0.02 pH; ± 0.5 mV
Calibration
Automatic at 1, 2 or 3 points with automatic recognition of 5 standard buffers
(pH 4.01, 6.86, 7.01, 9.18, 10.01) and 1 custom buffer
ORP
Range
± 2000.0 mV
Resolution
0.1 mV
Accuracy
± 1.0 mV
Calibration
Automatic at 1 custom point (relative mV)
DISSOLVED OXYGEN (HI98194, HI98196 only)
Range
0.0 to 500.0 %; 0.00 to 50.00 ppm (mg/L)
Resolution
0.1 %; 0.01 ppm (mg/L)
Accuracy
0.0 to 300.0 %:
± 1.5 % of reading or ± 1.0 % whichever is greater;
300.0 to 500.0 %:
± 3 % of reading
0.00 to 30.00 ppm (mg/L):
± 1.5 % of reading or ±0.10 ppm (mg/L)
whichever is greater;
30.00 ppm (mg/L) to 50.00 ppm (mg/L):
± 3 % of reading
Calibration
Automatic 1 or 2 points at 0, 100 % or 1 custom point