12
CONDUCTIVITY (HI98194, HI98195 only)
Range
0 to 200 mS/cm (absolute EC up to 400 mS/cm)
Resolution
Manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm
Automatic: 1 µS/cm from 0 to 9999 µS/cm;
0.01 mS/cm from 10.00 to 99.99 mS/cm;
0.1 mS/cm from 100.0 to 400.0 mS/cm
Automatic (mS/cm): 0.001 mS/cm from 0.000 to 9.999 mS/cm;
0.01 mS/cm from 10.00 to 99.99 mS/cm;
0.1 mS/cm from 100.0 to 400.0 mS/cm
Accuracy
±1 % of reading or ±1 µS/cm whichever is greater
Calibration
Automatic single point, with 6 standard solutions (84 µS/cm, 1413 µS/cm,
5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point
RESISTIVITY (HI98194, HI98195 only)
Range
0 to 999999 Ω·cm; 0 to 1000.0 kΩ·cm; 0 to 1.0000 MΩ·cm
Resolution
Depending on resistivity reading
Calibration
Based on conductivity or salinity calibration
TDS (Total Dissolved Solids) (HI98194, HI98195 only)
Range
0 to 400000 ppm (mg/L); (the maximum value depends on the TDS factor)
Resolution
Manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L);
1 ppt (g/L)
Automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L);
0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L);
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L);
Automatic ppt (g/L): 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L);
0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L);
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
Accuracy
±1 % of reading or ±1 ppm (mg/L) whichever is greater
Calibration
Based on conductivity or salinity calibration
SYSTEM SPECIFICA
TIONS
Summary of Contents for HI98194
Page 9: ...9 SENSOR AND PROBE INSTALLATION ...
Page 25: ...25 SETUP MENU STRUCTURE ...