35
Procedure
• From the “Calibration” menu select “Single param.
calibration” and then “ORP calibration”. The display shows
two options: “Custom ORP” and “Restore factory calib.”.
• For a user calibration select “Custom ORP”.
• Fill a beaker with an ORP test solution (see APPENDIX C
“Accessories”).
• Using the keypad, insert the numerical ORP value and then
press
Accept
to confirm.
• The stability counter will count down and the message
“Ready” and
Confirm
will be displayed.
• Press
Confirm
to accept the calibration point.
• After confirmation, the following messages are displayed:
“Storing” and “Calibration completed”.
• Press
OK
to return to the Calibration menu.
• Press
Measure
to return to the measurement screen.
• To restore the factory calibration data, select the corresponding option in the “ORP calibration”
menu and then press
Select
.
DISSOLVED OXYGEN CALIBRATION (HI98194, HI98196 only)
The accuracy of dissolved oxygen measurements is directly related to membrane cleanliness and
calibration technique. Oily coating and biological contaminants are the primary cause of calibration
drift in dissolved oxygen sensors.
Unfortunately, brushes or other cleaning objects may damage the membrane.
Replacing the membrane cap and electrolyte is the best way to perform periodic maintenance.
Although it may be easier to calibrate the D.O. sensor prior to deployment, it is advised to calibrate
at the site of deployment. Errors in measurement may result if altitude and barometric pressure differ
between the calibration and measurement site.
Note: Perform either the % D.O. saturation or D.O. concentration calibration.
If the % D.O. saturation range is calibrated, the D.O.
concentration range will also be calibrated, and vice
versa. Dissolved oxygen concentration values are based on
% D.O. saturation, temperature, salinity and atmospheric pres‑
sure. A standard solution or a reference D.O. meter may be used
to compare readings during calibration.
RELA
TIVE mV CALIBRA
TION
D.O
. CALIBRA
TION (HI98194, HI98196 only)
Summary of Contents for HI98194
Page 9: ...9 SENSOR AND PROBE INSTALLATION ...
Page 25: ...25 SETUP MENU STRUCTURE ...