• Immerse the sensor into the rinse standard by raising and lowering the beaker a few times to
ensure that the EC sensor channels are filled with fresh standard.
• Place the calibration beaker over the EC sensor and dislodge any trapped bubbles. Screw the
beaker into place. Wait for the reading to stabilize.
• The main display shows the actual reading, while the secondary level displays the current
temperature and the standard value.
• To change the standard value, press
Cal. point
and the list of available standard values is
displayed: 0 µS/cm, 84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and
111.8 mS/cm.
• The third level displays the status message.
• Press
Custom
to insert a custom value (temperature compensated value). Insert the desired value
using the keypad, then press
Accept
.
• When the reading becomes stable, the stability timer will count down and
Confirm
will appear.
Press
Confirm
to save the calibration.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press
OK
to return to the “Calibration” menu.
• Press
ESC
twice to return to main menu.
• Press
Measure
to return to the measurement screen.
Absolute Conductivity
• Select “Absolute conductivity” from the “Conductivity calibration” menu.
• Use the keypad to enter the custom value with the desired
resolution. Press
Accept
to confirm.
• Fill the calibration beaker with conductivity standard with
known conductivity at the temperature of standardization.
• Pour additional standard into a second beaker to be used to
rinse the sensor.
• Immerse the sensor into the rinse beaker and raise and lower the beaker to ensure that the EC
sensor channels are filled with fresh standard.
CONDUCTIVITY CALIBRA
TION (HI98194, HI98195 only)