42
Place
HI9819X
in a wind‑free area and choose “Custom
pressure” to perform a user calibration or “Restore factory calib”.
Note: “Custom pressure” procedure requires a reference
barometer.
Select the “Atm. pressure” from the “Calibration” menu.
• Select the “Custom pressure” option.
• Using the keypad, insert the numeric value that agrees with
the reference meter and then press
Accept
to confirm.
• The stability counter will count down and the message
“Ready” and “Confirm” will be displayed. Press
Confirm
to
store the calibration point.
• After confirmation, the following messages are displayed:
“Storing” and “Calibration completed”.
• Press
Measure
to return to the measurement screen.
• Press
OK
to return to the “Calibration” menu.
• To restore the factory calibration, select “Restore factory calib.” in the “Pressure calibration” menu
and press
Select
.
ATMOSPHERIC PRESSURE CALIBRA
TION
Chapter 8 - SYSTEM SETUP
From the main menu, select “System setup” and then “Meter
setup” or “Probe setup”.
Note: If the password protection is enabled, you will
be required to enter the password before any settings
can be modified.
Time
The meter uses a real time clock for logging. The time and time
format are set in this function.
Press
Modify
and set the time using the keypad.
Press
Accept
to save the time. When using the 12 hour format,
press A or P on the keypad for AM or PM after you set the time.
Press
Format
to change between 12 and 24 hour formats.
The default format is 24 hours.