18
PARAMETER SETUP MENU
7.4. AVERAGING
From the Parameter setup menu, use the
/
keys to highlight
Averaging
and then press
Select
.
Option: 1 to 20 readings
Press
Modify
to enter the
Averaging
screen. Use the alphanumeric keypad to enter the value, press
Accept
to save or
ESC
to return to the menu.
Averaging is a software filter used to minimize sensor noises and provide a more stable reading.
Averaging is particularly useful to get a representative reading from flowing water. Averaging will
affect all measurements. Readings are taken every second. During logging, the first sample will be
delayed by a few seconds if averaging is used.