27
CALIBRA
TION
RESTORING FACTORY CALIBRATION
Restoring factory calibration will delete the user calibration and restore the factory default calibration.
Use the
/
keys to highlight
Restore factory calib.
and then press
Select
.
8.5. ATMOSPHERIC PRESSURE CALIBRATION
The
HI98199
is factory calibrated for atmosphere pressure. The user can perform a single point
calibration or restore the factory calibration.
From the
Calibration
menu, use the
/
keys to highlight
Atm. Pressure
and then press
Select
.
• Select the
Custom pressure
option.
• Use the alphanumeric keypad to enter the known value from the reference meter and then press
Accept
to confirm.
• The stability counter will count down and the message “Ready” and
Confirm
will be displayed.
Press
Confirm
to store the calibration point.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press
Measure
to return to the measurement screen.
• Press
OK
to return to the
Calibration
menu.
RESTORING FACTORY CALIBRATION
Restoring factory calibration will delete the user calibration and restore the factory default calibration.
Use the
/
keys to highlight
Restore factory calib.
and then press
Select
.