26
27
• Press
to exit custom buffer edit menu. If the Saving Confirmation is enabled, press
to accept
the modified option,
to escape without saving or
to return to the editing mode. Otherwise,
the modified option is saved automatically.
• Use
key to select the next custom buffer to be set or press
to return to Calibration options.
Edit Buffer Group
Accessing this option, the user can edit the desired group of five pH
buffers for automatic buffer recognition (Automatic Buffer Entry
Type). If the Buffer Group already contains five pH buffers, at least
one pH buffer has to be removed in order to add another buffer.
To edit/set the Buffer Group:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Calibration option.
• Press
and use
or
to highlight the Edit
Buffer Group option.
• Press
and use
and
to choose the pH buffer to be included in the buffer group.
• Press
or
to add/remove the selected pH buffer to/from the buffer group.
• Press
to return to Calibration options and to save the changes.
Calibration Reminder
This option allows the user to select a calibration reminder
schedule. Three options are available for the calibration reminder:
Daily, Periodic or Disabled.
To set the Calibration Reminder:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Calibration option.
• Press
and use
or
to highlight the
Calibration Reminder option.
• Press
and use
or
to highlight the desired option.
• Press
to confirm your selection or press
to cancel operation.
Set Reminder Period
Schedule the calibration reminder timing with this option (verify
Daily or Periodic is set for Calibration Reminder).
If a Daily reminder is desired, set the time of day you wish the
reminder to occur.
If a Periodic reminder is desired, schedule time in days, hours and/
or minutes after the last calibration for the reminder to occur.
To set the Reminder Period:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Calibration option.
• Press
and use
or
to highlight the Set Reminder Period option.
• Press
and use
/
to select next/previous entry to be edited.
• Press
and use
or
to set the desired value, then press
to save the modified
value.
• Press
to return to the Calibration options. If the Saving Confirmation is enabled, press
to
accept the modified option,
to escape without saving or
to return to the editing mode.
Otherwise, the modified option is saved automatically.
Clear Calibration
This feature deletes the pH electrode calibration. A default pH
calibration will replace the actual electrode calibration until a new
electrode calibration is made.
To clear Calibration:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Calibration option.
• Press
and use
or
to highlight the
Clear Calibration option.
• Press
to clear calibration. A pop-up menu will be
displayed asking for confirmation (when a calibration is available).
• Press
to confirm or press
to escape without saving and return to the Calibration options.