28
29
Sample ID
This option allows the user to assign an identification number/
name. Two Sample ID options are available: ID Increment and Edit
Sample ID.
ID Increment
Two choices are available for the sample ID:
None
- the sample ID will be fixed and it can be entered
alphanumerically (see Edit Sample ID).
Automatic
- the sample ID will automatically increment by one
for each new log lot.
To set the ID Increment mode:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Sample ID option.
• Press
and use
or
to highlight the ID Increment option.
• Press
/
as desired.
• Press
to return to previous menu.
Edit Sample ID
This option allows the user to edit the sample ID.
Note
: The ID Increment mode must be set to None, to use this
feature.
To edit the Sample ID:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Sample ID option.
• Press
and use
or
to highlight the Edit
Sample ID option.
• Press
to confirm your selection.
• If the selected increment is None, the Text Editor menu will be displayed on the LCD, allowing you to enter the
desired sample number/name by accepting the highlighted character which is added in the text bar, using
. The
and
keys help the user to select the desired character.
It is also possible to delete the last character; position the cursor on the Backspace character and press
.
• Press
to return to Sample ID options. If the Saving
Confirmation is enabled, press
to accept the modified
option,
to escape without saving or
to return to
the editing mode. Otherwise, the modified options are saved
automatically.
• If the selected increment mode is Automatic, the desired sample
ID value can be set using
or
.
• Press
to save the current value or press
to
cancel operation.
Stability Criteria
This option allows the user to select the signal stability criterion for the measured parameter (pH, mV, ISE):
Fast
- this setting will give faster results with less accuracy.
Medium
- this setting will give medium speed results with medium accuracy.
Accurate
- this setting will give slower results with high accuracy.
To set the Stability Criteria:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Stability Criteria option.
• Press
and use
or
to highlight the
desired option.
• Press
to confirm your selection or press
to
cancel operation.
Reading Mode
This option allows the user to select between Direct and Direct/AutoHold pH mode.
Direct - the current reading is displayed in realtime on the LCD.
Direct/AutoHold - the current reading can be frozen on the LCD when
is pressed and the stability criterion is reached.