34
35
mV SETUP
pH Resolution
Select the desired pH resolution with this option. Choose from
one (X.X), two (X.XX) or three (X.XXX) digits displayed past the
decimals.
To set the pH Resolution:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the pH Resolution option.
• Press
and use
or
to highlight the
desired option.
• Press
to confirm your selection or press
to cancel operation.
The mV Setup menu allows the user to set the parameters associated with mV and Relative mV
measurements.
Accessing mV Setup
• Press
while in
Measure
mode and then
or
to select mV / Rel mV range for the desired
channel.
• Press
and then
to access mV Setup menu.
To access a mV Setup option:
• Use
or
to highlight the desired option.
• Press
to access the selected option.
The following is a detailed description of the mV Setup option
screens.
Profile
See pH Setup section.
Temperature
ORP measurements are not temperature compensated, although
ORP values can change with temperature (e.g. reference electrode
potential changes, sample equilibrium changes). It is important to
report ORP values together with the reference electrode used and
the temperature of measurement.
This option permits selection of the temperature source and
measurement units.