94
95
3) Direct/AutoHold Reading Mode and Automatic Log
Press
and then
keys must be pressed on
front display to initiate this function. Real time continuous
measurements are on display with “
AutoHold
” flashing and
real time continuous logging into meter memory, until the meter
reaches the stability criteria to go into Auto Hold mode. The
stored sample logs will be marked with an
“H”
to indicate the
Auto Hold mode. The virtual key
returns operation to
real time continuous measurements and
stops the logging
session.
4) Direct/AutoHold Reading Mode and Manual Log
Press
in order to add one new record in the log report. The manual log is working even if it is in
Auto Hold or Continuous reading mode. Press
to initiate the Auto Hold event. “
AutoHold
” will
flash until the stability criteria is reached and then the screen freezes in Auto Hold mode, the data is marked
with an
“H”
.
5) Direct/AutoHold Reading Mode and Auto Hold Log
Press
and then
keys to initiate and automate the capture of stable data which is stored in
the Recall Manual Log file. During the process, “
AutoHold
” will flash until the stability criteria is reached
and then the screen freezes in Auto Hold mode, the data is logged and marked with an
“H”
. The virtual key
returns operation to Real time continuous measurement. Press
again to log a second stable
data point. The lot ID along with the record index will appear for short time on the top/left corner on the
selected channel window, every time a record will be added to the lot.
LOG RECALL
This feature allows the user to view all stored data. If no data
were logged, the “
No records were found.
” message will be
displayed on the LCD in the Log Recall screen. Otherwise, the
instrument will display all the memorized lots in accordance
with the selected option: Automatic Log, Manual Log or ISE
Method Report (
HI 5522
only) for Channel 1, or Automatic Log,
Manual Log or USP Reports for Channel 2.
To view the memorized data:
• Press
while in
Measure
mode.
• Press
. Choose channel and then select the log report type.
LOGGING
e
d
o
M
g
n
i
d
a
e
R
g
o
L
ll
a
c
e
R
g
o
l
)
1
(
ci
t
a
m
o
t
u
A
g
o
L
ci
t
a
m
o
t
u
A
tc
e
ri
D
)
2
(
l
a
u
n
a
M
g
o
L
l
a
u
n
a
M
)
A
N
(
d
lo
H
o
t
u
A
e
l
b
a
ci
l
p
p
A
t
o
N
)
3
(
ci
t
a
m
o
t
u
A
g
o
L
ci
t
a
m
o
t
u
A
d
l
o
H
o
t
u
A
/
tc
e
ri
D
)
4
(
l
a
u
n
a
M
g
o
L
l
a
u
n
a
M
)
5
(
d
lo
H
o
t
u
A
g
o
L
l
a
u
n
a
M
There are 5 ways the Reading Mode and Log may be configured together. The table below shows the
combinations and indicates where the completed log will be stored.
2) Direct Reading Mode and Manual Log:
Real time continuous measurements are on display and snapshots
of measurement data are stored in the Manual log when the user
presses
. Subsequent snapshots will be added to the
same Manual Lot every time the
is depressed unless
New Lot
is selected under
Log
options.
Note
: When the
is pressed the lot ID along with the
current record number will appear for short time on the
selected channel window on the top/left corner (e.g.
L033_MV 8 - this means lot ID L033_mV and recod
number 8).
1) Direct Reading Mode and Automatic Log:
Real time continuous measurements are on display and continuous
logs to meter memory. These are sometimes referred as interval
logs. Press
.
w
w
w
.
.c
om
in
fo
rm
at
io
n@
itm
.c
om
1.
80
0.
56
1.
81
87