17
Section 1: Measurement Principles and Hardware
© 2017 Hansatech Instruments Ltd
SNAGGING THE THERMOCOUPLE WHEN SAMPLES ARE BEING INSERTED OR
REMOVED.
The Dark-Adaptation Leaf-Clip and Fibre-Optic Adapter
A separate leaf-clip system has been developed for situations where ambient light is to be
excluded from the sample during measurement. This is suitable for experiments requiring dark-
adapted measurements e.g. screening applications measuring Fv/Fm or situations which require
adaptation of tissue to standardised doses of actinic light.
The first step in the process is to attach the fibre-optic adapter (Figure 1-8) to the sample end of
the fiber optic. This is done by sliding the fibre-optic (A, Figure 1-8) into the adapter mounting
collar (B, Figure 1-8) and securing it in position by tightening the side mounted retaining screw (C,
Figure 1-8). The adapter / fibre-optic assembly can then be sealed into the top of the dark-
adaptation clip (D, Figure 1-8) and the position of the fibre-optic fine-tuned by unthreading the
retaining screw in the adapter collar and sliding the fibre-optic into the desired position before
locking it by re-tightening the collar screw.
Fig. 1-8.
The dark-adaptation leaf-clip and fibre-optic adapter. The fibre-optic (A) is
held in the adapter mounting collar (B) by the side-mounted retaining screw (C). The
adapter / fibre combination locate onto a dark-adaptation clip (D). The sample (E)
is supported on a foam pad and exposed to the FMS when the sliding shutter (F) is
opened.
THE GEOMETRY OF THE FIBRE-OPTIC WILL DETERMINE HOW EVENLY THE
SAMPLE WILL BE ILLUMINATED AND THE STRENGTH OF FLUORESCENCE
SIGNAL RECEIVED. IT MAYBE USEFUL TO MAKE SEVERAL TRIAL
MEASUREMENTS WITH SAMPLE IN ORDER TO OPTIMISE THE SYSTEM FOR
EACH PARTICULAR APPLICATION.
Summary of Contents for FMS1
Page 1: ......
Page 7: ...Section 1 Measurement Principles and Hardware ...
Page 21: ...Section 2 Software Installation System Operation Data Handling ...
Page 70: ...70 FMS 1 Fluorescence Monitoring System 2017 Hansatech Instruments Ltd ...
Page 71: ...Section 3 Stand Alone Instrument Operation and Data Upload ...
Page 83: ...83 Section 3 Stand Alone Instrument Operation and Data Upload 2017 Hansatech Instruments Ltd ...