DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
59
[Clock Source]
Select CH1
~
CH4 or LA as the trigger source.
Note: LA must be inserted when LA is used as the trigger source.
[Slope]
Select the required triggering edge (rising edge, falling edge, double edge) and
press V0 to confirm.
[Data source]
Select CH1
~
CH4 or LA as the trigger source.
Note: LA must be inserted when LA is used as the trigger source.
[Overtime]
The timeout period must be greater than the period of the CLK clock source.
[Data bit width]
Set the data bit width.
[Data Bit]
Set the number of data bits.
[Mode]
Select the acquisition mode (auto, normal) and press V0 to confirm.
[Holdoff]
Set the holdoff time.
[Decode]
Switch the decoding function.
[Label position]
The position where the decoded data is drawn on the screen (only
available when the decode function is turned on).
[Table]
Data recording mode.
Trigger level knob: The analog channel can modify the trigger level value. Trigger the
mark and move up and down with the knob.
Digital channel: You can change the trigger threshold of the digital channel by setting
the threshold voltage.
Trigger settings:
Clock source: CH3; Slope: Rise; Data source: CH4; Overtime: 366ns; Data bit width: 8;
Data bit: 7 bits. Adjust the trigger level. The trigger result is displayed as follows: