5. Positive pulse width [+ Width]: The time difference from the middle value of the
threshold of the rising edge of the pulse to the middle value of the threshold of the next
falling edge.
6. Negative pulse width [-Width]: The time difference from the threshold intermediate
value of the falling edge of the pulse to the threshold intermediate value of the next ris-
ing edge.
7. Positive Duty Cycle [+ Duty]: The ratio of positive pulse width to period.
+Duty =
positive pulse
period
× 100%
8. Negative Duty Cycle [-Duty]: The ratio of negative pulse width to period.
−Duty =
negative pulse
period
× 100%
9. Burst Length [B-Width]: The time from the first edge of the data source to the last
edge of the data source.
10. Maximum time [V
max
Time]: The time value corresponding to the maximum value
(V
max
) of the waveform.
11. Minimum time [V
min
Time]: The time value corresponding to the minimum value (V
min
)
of the waveform.
Note: The upper threshold, middle threshold, and lower threshold are 90%, 50%, and
10%, respectively.
Count value parameters
1. Positive Pulse Count [+ Pulse Count]: The number of positive pulses rising from be-
low the lower threshold to above the upper threshold.
Threshold upper limit
Threshold lower limit
1
2
3
n
n-1
2. Negative pulse number [- Pulse Count]: the number of negative pulses from above
the upper threshold to below the lower threshold.