DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
72
The calculation formula of the phase is:
Phase =
𝐷𝑒𝑙𝑎𝑦
𝑝𝑒𝑟𝑖𝑜𝑑1
× 360
°
Among them,
Phase means "phase ↑ A-> B" or "phase ↓ A-> B"
Delay means "Delay FRR" or "Delay FFF"
Period1 represents the period of digital data source source 1.
Voltage parameters
Threshold upper limit
Threshold Mid- limit
Threshold lower limit
Overshoot
Amplitude
Vmax
Vtop
Vbase
Vmin
Pk-Pk
Preshoot
1. Maximum [V
max
]: The voltage value from the highest point of the waveform to GND.
2. Minimum [V
min
]: The voltage value from the lowest point of the waveform to GND.
3. Double peak [P
k
-
P
k
]: The voltage value from the highest point to the lowest point of
the waveform.
Double peak = Maximum-Minimum
4. Top value [V
top
]: The voltage value from the flat top of the waveform to GND.
5. Bottom value [V
base
]: The voltage value from the flat bottom of the waveform to GND.
6. Amplitude [V
ampl
]: The voltage value from the top to the bottom of the waveform.
Amplitude = Top value-Low value
7. Middle value [V
mid
]: The actual voltage value corresponding to the middle value of
the measurement threshold.
8. Average: Arithmetic average over the entire waveform or gated area.
9. Period average: The arithmetic average of the first period of the waveform.