DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
95
Start the upgrade process.
3. Use multi-function knob V0 to select the .upk file to be upgraded, and press multi-
function knob V0 to select the upgrade file.
4. After the upgrade is completed, a prompt message will pop up on the screen to re-
mind the user that the upgrade has been successful and the oscilloscope will restart
automatically.
5. Press the F5 back key to cancel the upgrade.
Calibration
Performing the calibration procedure can quickly make the oscilloscope reach the best
working condition to obtain the most accurate measurement value. The user can exe-
cute the program at any time, especially when the ambient temperature variation range
reaches or exceeds 5 ° C. Before performing the self-calibration operation, make sure
that the oscilloscope has been warmed up or run for more than 30 minutes. Disconnect
all input channels and press Utility-> Calibration.
AUX
Press Utility -> AUX to set the type of signal output from the Aux port on the rear panel.
Trigger output
The oscilloscope generates a trigger, and the interface will output a signal reflecting the
current capture rate of the oscilloscope. Connect this port to the waveform display de-
vice to measure the frequency of the output signal. The measurement result is the
same as the current capture rate.
Pass / fail
When the oscilloscope tests a waveform that meets the pass / fail rule and output mode,
it will output a pulse signal. For details, please refer to the introduction in "
System settings
Sound
Press Utility-> System Settings-> Sound. After turning on the sound, you can hear the
buzzer sound when operating the menu or popping up a prompt message. Sound is
turned on by default.