DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
88
Run the [Pass / Fail] test, turn on the "Output" and select "Output to DDS", the system
will automatically turn on the "Burst" switch under the corresponding signal source, and
the "Count" defaults to 1. According to the test results, the specified signal source out-
puts a waveform with a set number of cycles.
DSO
Trigger
:
Each time the waveform in the oscilloscope triggers, the current signal source outputs a
waveform with a set number of cycles.
Pass / Fail test
The pass / fail test function is mainly used to evaluate the signal quality. By setting rules
in the oscilloscope to test the waveforms in real time, counting the number of tests with-
in the specified test conditions, the number of pass / fail data frames, and displaying
the test results.
Enable pass / fail test
Press [Utility] -> "Pass / Fail"-> “Enable test”->"On" to enable the test function. The
shaded area indicates the failure area. If the waveform measured at a time during the
measurement passes through the failure area, it is determined as a failure frame.
Note: When the Time Mode is set to X-Y mode, ROLL mode or the horizontal time base
is set to 200 ms / div or slower in YT mode, when the instrument enters the "slow scan"
mode, this function cannot be enabled.
1. Data source
Before selecting the data source, you need to connect the signal under test to the ana-
log channel input of the oscilloscope. Press the "Data Source" soft key to select the
channel (CH1-CH4) to be tested. Note: Only channels that are already open can be se-
lected.
2. Rule
Users can customize the rules for pass / fail tests. Press the "Horizontal" and "Vertical"
soft keys respectively, rotate V0 to adjust the pass / fail rule.
Press the "Create" soft key
to apply the current rule. The oscilloscope prompts "Mask created!".
3. Test and output
The user can set the output format of the test results as follows.
Press the
Message
soft key to select whether to display the test results.
Press the
Stop On Output
soft key and select On or Off.
On: When the oscilloscope detects a waveform that matches the set “Mode”, the test
stops immediately and enters the STOP state. The information displayed on the screen
at this time is the statistical result as of STOP. If “On” is selected for “Output”, the pulse