background image

The Hantek HT300B 300 MHz and HT500B 500 MHz passive 

probes are high impedance probes with switchable attenuation 

ratio (1:1 and 10:1). These are compatible with  hantek digital 

storage oscilloscopes or similar oscilloscopes with 1 Mohm 

BNC input. 

 Table 3 - Physical & Environmental Characteristics 

Characteristic

Value

Probe Body

Cable Length
Weight

95mm
1200mm
55g

 Table 2 - Electrical Characteristics 

Characteristic 

Value

Attenuation Ratio

Input Capacitance
Compensation range
(when terminated 
into 1 Mohm)

1:1 / 10:1

1:1 - 100pf

15pf - 40pf

Input Resistance

Maximum Input 
Voltage

1:1  - 150V rms CAT II

1:1 - 1M

Ω

 

Temperature

Humidity

Altitude

The BNC connector of these probes has a 
pin hole without a pin. This is as per the 
design and does not indicate any damage 
or fault in the probe.

BNC 
Connector

Pin hole 
without 
a pin

These probes have been designed and calibrated for use 
with instruments having an input impedance

 of 1M Ω

 

paralleled by 20pF. However, these may be recalibrated for 
use with instruments having an input capacitance of 
15 to 40pF.

Adjusting Probes for Low-Frequency (LF) Compensation

The probe should be adjusted for LF compensation when it is 
connected to an oscilloscope input for the first time. LF 
compensation matches the probe cable capacitance to the 

oscilloscope input capacitance. This matching assures good 
amplitude accuracy from DC to the upper bandwidth limit 
frequencies. A poorly compensated probe clearly influences 
the overall system performance (probe and oscilloscope) 
and introduces measurement errors resulting in inaccurate 
readings and distorted waveforms. 

To perform LF compensation:

1 Connect the probe to the oscilloscope’s front-panel 
   calibration output (a square wave label is usually seen near 
   this output).
2 Use the supplied adjustment tool to adjust the LF 
   compensation to an optimum square wave response as 
   shown in the figure below.

10:1 - 15pf

10:1 - 10M

Ω

 Table 1 - Probe and its Supplied Accessories 

 ID Part

Quantity

For 

HT500B

For 

HT300B

Hook Tip

Ground Spring
Ground Lead

Adjustment Tool

Protective Cap
Channel 
Identification rings

2
2
2

2

-

2

1

1

1

1

  2 sets of each color with 
  two rings in each set

Undercompensated

Overcompensated

Optimum

10:1  - 300V rms CAT II

3,000 m (9,842 ft) (Operating)
12200 m (40,026 ft) (Non-operating)

Bandwidth

For 

HT300B

 

1:1  -  6  MHz

 

10:1 - 

35

0 MHz

For 

HT500B

1:1 - 6 MHz
10:1 - 

500

 MHz

5% to 95% RH up to 30 °C  (Operating)
5% to 65% RH above 30 °C up to 55 °C 

-10 °C to +55 °C (Operating)

-40 °C to +70 °C (Non-operating)

Passive probe

2