background image

 

 

 

Differential 
overvoltage 
detection level 

1/50: 140V ± 4V (DC + peak AC) 
1/500: 1400V ± 40V (DC + peak AC) 

AC adapter output 

DC9V

1000mA 

AC adapter input 

AC100~240V

50~60Hz 

Typical mechanical specification 

Input cable length 

HT8050

60cm

HT8100

30cm 

Output cable length  90cm 
Alligator clips 

85×40×13

mm

 

Pincer clips 

160×45×13

mm

 

Probe body 

120×55×24

mm

 

Weight (probe only)  160g 

Environment Specification 

Operating 
temperature 

0~40

 

Storing temperature  -10~45

 

Operating humidity  85%RH 
Storing humidity 

90%RH 

Operating altitude 

3000m 

Storing altitude 

12000m 

 

 

 

 

Summary of Contents for HT8000 Series

Page 1: ...HT8000 Series High Voltage Isolating Probes Instruction Manual Qingdao Hantek Electonics Co Ltd...

Page 2: ...1 HT8000 Series High Voltage Isolating Probes Instruction Manual Please read the instruction manual carefully to avoid injury and prevent damage to this product Ver 2 01 2022 11 16...

Page 3: ...a replacement in exchange for the defective product Parts modules and replacement products used by Hantek for warranty work may be new or reconditioned to like new performance All replaced parts modu...

Page 4: ...able of contents 1 Safety Summary 4 2 Brief Introduction 5 3 Main Specification 6 4 Unpacking and Preparation for Use 8 5 General View of Product 9 6 General View of Accessories 10 7 Operating Basics...

Page 5: ...sconnect probes while they are connected to a voltage source Ground the product This product is indirectly grounded through the grounding conductor of the mainframe power cord To avoid electric shock...

Page 6: ...C peak AC accordingly HT8000 Series high voltage isolating probe has over range alarm function It indicates with a red LED HT8000 Series high voltage isolating probe is small and robust It has very hi...

Page 7: ...ce Between each input to ground 4M Between inputs 8M Input capacitance Between each input to ground 7pF Between inputs 3 5pF Maximum measurable differential voltage 1 50 130V DC peak AC 1 500 1300V DC...

Page 8: ...echanical specification Input cable length HT8050 60cm HT8100 30cm Output cable length 90cm Alligator clips 85 40 13 mm Pincer clips 160 45 13 mm Probe body 120 55 24 mm Weight probe only 160g Environ...

Page 9: ...Instruction manual 1pcs AC adapter 1pcs Red alligator clip 1pcs Black alligator clip 1pcs Red pincer clip 1pcs Black pincer clip 1pcs 4 2 Preparation for use Please check the line voltage prior to con...

Page 10: ...9 5 General View of Product...

Page 11: ...View of Accessories HT8000 Series high voltage isolating probe includes following accessories An AC adapter a couple of alligator clips and a couple of pincer clips 1 AC adapter 2 Alligator clips 3 P...

Page 12: ...probe power input connector The green POWER indicator lights e Select adapt attenuation range according to measured voltage When the measured voltage is over range the red OVER RANGE indicator lights...

Page 13: ...ld be lower than 10MHz Otherwise the errors will be larger c The probe and measuring instrument should warm up for 20 minutes if want to get high accuracy measuring result 7 3 Troubleshooting The Powe...