Quick start
EN
Tablet1000 User manua
l
Copyright © Qingdao Hantek Electronics Co., LTD
14
Figure 4.3 Probe41
Manual probe compensation
When the probe is connected to any input channel for the first time, perform this
adjustment to match the probe to the input channel. Probes that are not compensated or
compensated with bias can lead to measurement errors. To adjust the probe
compensation, please follow the following steps:
1)
Connect one end of the probe to the CH1 channel and the other end to the AFG
channel.
2)
Enter the oscilloscope interface, click
Utility
->
Probe calibrate
, and set the switch
of the probe to 10X as prompted.
3)
Check the shape of the waveform displayed
Figure 4.4 Probe compensation4
3) If necessary, adjust the variable capacitance on the probe with a non-metallic
screwdriver until the waveform displayed on the screen is "compensated correctly" as
shown above.