BX176 portable single gas detector operation manual
19
9. Trouble shooting guidance
Failure
Possible failure cause
Solution
Cannot power on
Low voltage
Charge the detector in time
System halted
Contact to the seller
Circuit fault
Contact to the seller
No response to gas
response time
incomplete
Wait for sensor response
Circuit fault
Contact to the seller
Inaccurate detect
Sensor overdue
Replace the sensor
Long time no
calibration
Recalibrate the sensor
Time error
Battery voltage
exhausted
Charge the detector in time
and set the time again
Electro Magnetic
Interference
Set the time again
Zero translation function
unavailable
Sensor drift too much
Calibrate the sensor or replace
the sensor in time
Detecting mode show“-0”
Sensor drift
Zero translation
Detecting mode show as
below picture:
Sensor fault
Replace the sensor