Chapter 3 - Installation
9
CHAPTER 3 - INSTALLATION
Precautions Do
not
store or operate the scale out of its
specified temperature range.
Do not
store other equipment on the scale when
it is not in use or in storage.
Do not
allow a build-up of debris on, around or
under the scale.
Do not
set the scale in water or allow water to
settle around the scale. Provide Proper drainage.
Do not
allow static or other electrical discharges
through the scale.
Do not
try to repair the scale. Contact Hardy
Process Solutions Technical Support Department
for all repairs.
Unpacking
Hardy Bench scales are shipped fully assembled
and only require adjustment of the overload
stops and adjustment of the leveling feet. If there
is any damage to the shipping container or to the
scale, save all the packaging material and file a
report with the shipper.
Adjusting the
Mechanical Overload
Stops
There is a bolt in each corner of the frame that is
fully engaged for shipping protection. Each
should be adjusted to allow the load cell to
deflect to its maximum deflection distance, but
no further.
Maximum Deflection at Capacity:
HISP6 - 0.17mm
HISPA22 - 0.4mm