1. Safety 

Intended Application 

PXI Express PXCe4006 series chassis are not end-products, so there is no valid approval for this unit. In 
Order to enable stand-alone functionality, additional elements are required. An operational system is 
achieved only by way of appropriate Measurement Cards. 

The completion and final testing of the units have been carried out, or at least supervised, by qualified 
technicians. These instructions are directed exclusively to these qualified technicians i.e. Engineers, 
trained and qualified electricians etc. 

Make sure that the finished system complies with the safety regulations currently applicable in the 
country it is going to be used. 

 

Safety Symbols 

 

 

Hazardous voltage! 

Familiarise yourself with the danger of electrical voltages and the safety 
precautions before starting to work with parts that carry dangerous voltages

 

 

 

 

Caution!

 

This symbol indicates a condition where damage of the equipment or injury of 
the service personnel could occur. To reduce the risk of damage or injury, 
follow all steps or procedures as instructed.

 

 

 

 

Danger of electrostatic discharge!

 

Static electricity can damage sensitive components in a system. To avoid 
damage, wear ESD wrist straps or at regular intervals touch blank enclosure 
parts. 

 

General Safety Precautions  

 

 

Warning!

 

Voltages over 60 VDC can be present in this equipment. This equipment is 
intended to be accessed, to be installed and maintained by qualified and trained 
service personnel only. 

This equipment is designed in accordance with protection class 1! 

lt must therefore be operated only with protective GND/earth connection! 

 

Summary of Contents for PXCe4006

Page 1: ...www hartmann electronic com Rev 1 0 1 PXIe Slimbox Chassis PXCe4006 6 Slot 1U 19 User Manual...

Page 2: ...Hartmann Electronic GmbH Motorstra e 43 D 70499 Stuttgart Weilimdorf Telephone 49 711 1 39 89 0 Telefax 49 711 8 66 11 91 E Mail info hartmann electronic de Internet www hartmann electronic com Hartm...

Page 3: ...lock 13 Front Panel LEDs 15 Interoperability with CompactPCI 15 3 Installation 16 Connecting Chassis Grounding 16 Inspecting the Chassis Components 16 Protection Against Electromagnetic Interference 1...

Page 4: ...bols Hazardous voltage Familiarise yourself with the danger of electrical voltages and the safety precautions before starting to work with parts that carry dangerous voltages Caution This symbol indic...

Page 5: ...rostatic Discharge Precautions Electronic components can easily be destroyed by electrostatic discharge which can occur between chassis components and a person Before working on the chassis make sure...

Page 6: ...is properly grounded at all times the following conditions shall be met This equipment shall be connected directly to the AC supply system earthing Board Installation Electrostatic discharge and incor...

Page 7: ...h PXI and PXI Express The key features of the PXCe4006 Series Slimbox include the following Optional Rear Panel PXI_TRIG 7 0 SMB I O Connector Rear Panel PXI_CLK10 I O Connector Removable Fan Tray to...

Page 8: ...com Rev 1 0 8 Chassis Description Figure 2 1 Front view of the PXCe4006 series Slimbox 1 System slot 5 Power LED 2 PXIe peripheral slot 6 On Standby momentary power switch 3 PXIe hybrid slots 7 Remov...

Page 9: ...ry switch 6 at the front Panel Position 2 Fan Control The Chassis cooling fans are running at full speed Fan control is enabled Fan Control The Chassis Cooling Fans are controlled through two temperat...

Page 10: ...www hartmann electronic com Rev 1 0 10...

Page 11: ...thout the momentary switch on the front panel Therefore the auto mode of the chassis has to be turned on The auto mode can be turned on by changing the DIP Switch s position 1 on the chassis rear pane...

Page 12: ...slot using PCIe X1 These slots have up to 500MB s single direction dedicated bandwidth X1 Gen 2 PCI Express PXI Local Bus The PXIe backplane local bus is a daisy chained bus that connects each periphe...

Page 13: ...s are as shown in Figure 2 6 The PXCe4006 Series Slimbox provides an external 10 MHz PXI_CLK10 SMB Clock input If a Clock signal is detected on the rear panel SMB connector the backplane automatically...

Page 14: ...clock if the input clock is present The presence of the input clock is automatically detected If there is no input clock present the clock source to the PLL is a 10 MHz Clock oscillator Each of the c...

Page 15: ...teroperability with CompactPCI The PXCe4006 series Slimbox allows you the flexibility to use the following devices in a single PXI Express chassis PXI Express compatible products CompactPCI Express co...

Page 16: ...s There is a 4 mm stud at the rear panel This stud is only for potential equalization Grounding is achieved through the protective earth conductor of the power cable Inspecting the Chassis Components...

Page 17: ...amage to the interface connector pins 5 Install filler panels in unused or empty slots Filler panels consists of a front panel with or w o air baffles EMC gasket and mounting screws WARNING Close all...

Page 18: ...e should be taken that NI PXI Platform Services is installed before the Hartmann Electronic installer is executed Make sure that the installer is executed within the folder that is delivered with the...

Page 19: ...the shelf until the interface connector disengages Wait until the fans have stopped and then remove the fan tray from the shelf WARNING Extreme care should be taken while handling the fan tray It is r...

Page 20: ...ing to EN 60950 1 HARTMANN PXCe4006 Series Slimbox are not end products with independent functionality according the EMC regulations therefore CE marking is not required Not before CPCI PXI Express bo...

Page 21: ...rated output Over Voltage protection on 5 V 12 V 3 3 V The continuous total output power is 120 W max The combined Power of 5 V and 3 3 V is 70W max Table 6 1 shows the characteristics of the PXCe400...

Page 22: ...m Accuracy 25 ppm Output Impedance 50 10 External PXI_CLK10 Input SMB Rear panel connector Frequency 10 MHz 25 ppm Input sine wave or square wave 3 5V RMS into 50 10 24 dBm Operating Environment Opera...

Page 23: ...tronic com Rev 1 0 23 Mechanical Slimbox weight ca 6 kg Figure 6 1 PXCe4006 series Slimbox Dimensions Front Dimensions are in Millimeter Figure 6 2 PXCe4006 series Slimbox Dimensions Side Dimensions a...

Page 24: ...www hartmann electronic com Rev 1 0 24 Figure 6 3 PXCe4006 series Slimbox Dimensions Bottom Dimensions are in Millimeter...

Page 25: ...he system make sure that all slots are populated with either boards or filler panels Chassis Material 1 0330 st12 03 ZE Humidity Operating 5 to 80 non condensing Storage 5 to 80 non condensing Tested...

Page 26: ...Assignment The Pin assignment applies with the PXI 5 PXI Express Hardware Specification and the CompactPCI PICMG 2 0 R 3 0 specification System Controller Slot 4 Link Configuration PXI Express Hybrid...

Page 27: ...5 1PETp0 1PETn0 GND 1PERp0 1PERn0 GND 1PETp1 1PETn1 GND 6 1PETp2 1PETn2 GND 1PERp2 1PERn2 GND 1PERp1 1PERn1 GND 7 1PETp3 1PETn3 GND 1PERp3 1PERn3 GND 2PETp0 2PETn0 GND 8 2PETp1 2PETn1 GND 2PERp1 2PER...

Page 28: ...ND RSV PXI_LBL6 PXI_LBR6 GND Pin A B ab C D cd E F ef XP3 Connector 1 PXIe_CLK100 PXIe_CLK100 GND PXIe_SYNC100 PXIe_SYNC100 GND NC NC GND 2 PRSNT PWREN GND NC NC GND NC NC GND 3 SMBDAT SMBCLK GND RSV...

Page 29: ...R C BE 1 GND 17 GND 3 3V IPMB_SCL IPMB_SDA GND PERR GND 16 GND DEVSEL GND V I O STOP LOCK GND 15 GND 3 3V FRAME IRDY BD_SEL TRDY GND 12 14 Key Area 11 GND AD 18 AD 17 AD 16 GND C BE 2 GND 10 GND AD 21...

Page 30: ...PXI_LBL6 PXI_LBR6 GND Pin A B ab C D cd E F ef XP3 Connector 1 PXIe_CLK100 PXIe_CLK100 GND PXIe_SYNC100 PXIe_SYNC100 GND NC NC GND 2 PRSNT PWREN GND NC NC GND NC NC GND 3 SMBDAT SMBCLK GND RSV RSV GND...

Page 31: ...www hartmann electronic com Rev 1 0 31 8 Ordering Information Ordering Number Description PCI V I O LMH0000792 PXI Express 6 Slot 1U Slimbox 5 V...