29
5.3 Waveform Generator (cont’d)
5.3.6 Software Maintenance Mode (cont’d)
5.3.6.3 Pulse Protocol
The pulse protocol to be delivered at run time is set by pressing
the PROTOCOL button and adjusting parameters via the
touch screen interface. Parameters for three pulse groups are
modifiable by selecting a box and scrolling through the available
values. Subsequent groups are activated by selecting more than
one pulse in the number field and conversely deactivated by
selecting zero pulses in the number field.
The adjusted protocol is saved to internal memory when the
HOME button is pressed. Minimum and maximum parameter
values are as follows:
Protocol Parameter Boundaries
Amplitude (volts) Duration (ms)
Interval (ms)
Pulse #
Pulse Groups
Minimum
50
0.050
0.200 (5 kHz)
1
1
Maximum
1000
10*
1000 (1 Hz)
10
3
* For Pulse Amplitudes greater than 300V, the maximum allowable Pulse Width is 1ms
** Pulse duty cycle is limited to 50%
Caution
IMPORTANT! The changes made on this screen will
only be permanently saved when the HOME button
is pressed to go back to the Welcome Screen. If
modifications are made and system power is turned off
while still viewing the screen, all changes will be lost.
Figure 5-15: Pulse Protocol Modification
System Operation
5