AgilePulse Waveform Generator
Pulsing
Resistance Pulsing
4 µs at 5 volts every second
Pulse Protocol Parameters
Parameters in a Group:
Pulse Width
50 µs to 1 ms 50 to 1000 volts
50 µs to 10 ms 50 to 300 volts
Pulse Current Trip
26 amps
Load Range
100 to 1500 ohms
Number of Pulses
1 to 10
Maximum Duty Cycle 50%
Interval
200 µs to 1 sec
(pulse start to pulse start)
Number of Groups
3
Pulse Measurement
Internal Digitizer
Levels
12 bit
Samples
Pulse width/8 minimum 100 µs
Data Stored Internally and on External USB Key
Data Types
Raw Data
DV<Date>.xml
Log Data
DV<Date>.txt
CSV Data
DV<Date>.csv
All data automatically stored in internal memory and may be
downloaded to an external USB Key.
Maximum Data Logs stored and retrievable from internal flash
memory > 20,000.
Front Panel
Computer:
Operating System
Windows
®
Mobile 6.0
Interface
Touch screen
Line/Mains Switch with illumination
Emergency Stop Button (resets computer to ready state)
Touch Screen
USB Ports
2
Electrode connector Fischer Series 4032
Back Panel
Power Entry
IEC 320
Ethernet
RJ45
(NOT CURRENTLY INSTALLED)
Electrical and Mechanical
AgilePulse Cabinet with Handle 16.5 cm x 30.5 cm x 40.6 cm
(W x H x L)
6.5 in x 12 in x 16 in
Weight
5.4 kg (12 lbs)
Operating temperature
10 to 40 oC
Mains Voltage
100 to 250 VAC
Fuse
5 Amp Slo-Blo
®
, 5 mm x 20 mm
Appendix A: AgilePulse
™
ID Specifications
A