No.:GC-JS-21 Version: 01/00
Effective date
:
2020-03-30
16
5.2 Introduction to the load of each program
5.2.1 Load type corresponding to each program
134℃
universal
quick
Non-wrapped load or wrapped solid, fabric and cavity load, Paper-plastic packaging or
double-layer packaging may not dry, should be used immediately after sterilization
Universal
Non-wrapped load or wrapped solid, fabric and cavity load
134℃ solid
Solid load without package, should be used immediately after sterilization
B-D/Helix test
B-D test device or PCD test device
Vacuum test
Without any loading
Drying
Dry the load when it is not dry after sterilization
5.2.2 Applicable maximum load table
Model
Maximum weight of instruments
Maximum weight of fabric loads
A50B
2.2kg
0.5kg /package
5.3 Maximum working time and maximum water consumption when each program is at maximum
load
Model
Loading
kg
134℃
quick
Universal 134℃ Universal
121℃ Universal
134℃ Solid
Longes
t time
min
Min
water
consump
tion
mL
Longest
time
min
Min
water
consump
tion
mL
Longest
time min
Min
water
consump
tion
mL
Longest
time
min
Min
water
consump
tion
mL
134
℃
universal quick
134
Preset
4
3
4
℃
134
universal
Preset
134
3
6
4
4
134
℃ solid
134
1
2
121
℃
universal
121
Preset
8
3
4
B-D/Helix test
134
3.5
3
4
Test program
Vacuum test
Vacuum to -80kPa
,
holding 15min
Test program
Drying
5
min
Preset drying