6 Troubleshooting
15
6.1
Replacing fuse
Damage to equipment!
An incorrect fuse in the discharging power pack may cause a defect.
This may result in a cable fire.
• Only use fuses of the specified type.
• Never use repaired fuses.
• Never bridge the fuse.
The unit type and the rated voltage are indicated on the nameplate.
1. Disconnect the discharging
power pack.
2. Determine and rectify the
reason for the fuse failure.
3. Use a screwdriver to
unscrew the fuse holder (C)
and remove it.
4. Replace the fuse and attach
the fuse holder again.
Only use the following fuse:
• 100 V = T500 mA L250; 5 x 20 mm
• 115 V = T500 mA L250; 5 x 20 mm
• 230 V = T250 mA L250; 5 x 20 mm