Operation
Page 7
HAUG
Static Meter
Ι
3.3
Change measurement distance
Press and hold the "function/on"-key (approximately 2 s) until "change
distance" will appear in the second row. The pre-selected distance in cm will
be displayed in the first row. Press the "function/on"-key once shortly to
change the measurement distance. With each short press of the
"function/on"-key the proposed distance will appear as follows:
2 cm
è
5 cm
è
10 cm
è
20 cm
è
1 cm
è
2 cm (and so on).
When you reach the right distance, please wait (do not press the
"function/on"- key) until the actual measurement value appears in the second
row. The new selected distance is now also shown in the first row. Please
consider this selected distance for the following measurements. The distance
will be measured from the electrode positioned 3 mm apart from the housing
on the modulator side. It is necessary to add 3 mm to any selected distance
starting from the housing of the unit.
Please note:
The measurement distance will change back to 2 cm after each switch off and
on of the unit!
3.4
Distance support
With any equipment you will get 2 distance supports with a length of 26 mm.
The distance supports can be screwed into the aluminium plate on the
modulator side, which will always provide the right measurement distance for
the standard setting of distance = 2 cm.
3.5
Display
The alphanumeric Liquid Crystal Display (LCD) consists of 2 rows for 12 digits
each. The measurement distance in cm will be shown in the first row, while
the voltage reading in volts is displayed in the second row.
When the level of 999 volts is reached, the display will switch automatically
into the kV-mode. The display of the voltage reading appears always in 3
digits, such as: 578 V; 3.85 kV; 24.0 kV and 274 kV respectively.
Select a greater distance if "overflow !" appears on screen!