6 Troubleshooting
13
6
Troubleshooting
Electric shock hazard!
The discharging power pack is operated electrically and generates a
high electric voltage. In the event of any faults, there is a risk of an
electric shock.
• Faults may only be eliminated by a trained and qualified
electrician.
NOTE:
If the error cannot be removed in this way, return the discharging
power pack and ionizing unit for checking to HAUG GmbH & Co. KG
(the address is provided on the back of the envelope).
Fault
Cause
Troubleshooting
Mains
switch
does not
illuminate.
Power failure
Inspect the power fuse
Defective fuse in
discharging power
pack.
Check the fuse of the discharging
power pack.
Ionisation
not
available
HV not available
Inspect the connections on the
discharging power pack.
Electrically
conductive
deposits on
ionizing unit.
Clean ionisation unit
Discharging power
pack is damaged.
Shut down the discharging power
pack immediately and secure it to
prevent unintentional restart.