6 Troubleshooting
14
6.1
Replacing fuse
Damage to equipment!
An incorrect fuse in the discharging power pack may cause a defect.
This may result in a cable fire.
• Only use fuses of the specified type.
• Never use repaired fuses.
• Never bridge the fuse.
The unit type and the rated voltage are indicated on the nameplate.
1. Disconnect discharging
power pack from supply.
2. Determine and remove the
cause for the blown fuse.
3. Detach the fuse holder
using a screwdriver and lift
out.
4. Replace fuse and reattach
fuse holder.
Use the following fuse only:
• 100 V = T500 mA L250; 5 x 20 mm
• 115 V = T500 mA L250; 5 x 20 mm
• 230 V = T250 mA L250; 5 x 20 mm