rev. 122817MH
©2017 HAUSMANN INDUSTRIES, INC.
Fig. 4
Fig. 5
Fig. 6
5
LOCKING CASTERS
You must lock brakes of all (4) casters by depress-
ing brake pedal down before patient use.
PIVOTING FOOT PLATE
You must adjust angle position before patient use.
There are two black spring loaded control handles.
Lift up and turn knobs to “open”. Adjust foot plate
into an approximate 90° right angle position (there
are adjustment holes on steel rod). Turn both knobs
to re-lock the foot plate into desired position as
shown. Turn both knobs to snap lock the ft plate into
desired pos as shown in fig 6 below.
NOTE: DO NOT OPERATE TILT IF FOOT PLATE
IS IN “OUT OF WAY” POSITION AS SHOWN IN
FIG 5.
Foot plate shown in 90° operational position.