11.3.5
Synchronizing a Master/Slave setup to external time sources
In a Master/Slave setup, all time information is recorded by the Master
mainframe only. Synchronization source selection on Slaves is disabled as
Slave mainframes are dedicated to follow the Master mainframe.
For the Master mainframe, the date and time are controlled by either the PC
(RTC) , PTP master, IRIG or GPS, depending on the synchronization source
selected. The source is selected in the Perception software.
In the
Perception work
area:
Figure 11.11:
Perception work area synchronization source
A
Settings tab
B
General group
C
Sync source column
To set the synchronization source in the Perception software, proceed as
follows:
1
Select the
Settings
sheet.
2
In the
Settings
sheet, go to the
General
group in the task pane and select
Mainframe
.
A list of available mainframes is displayed in the settings area.
3
Select the Master mainframe.
4
Double
-
click on the
Sync source
cell to open it for modification.
5
In the drop
-
down list that appears, select the synchronization source that
should be used.
For information on how to check synchronization, please refer to appendix
"Master/Slave synchronization verification procedure" on page 901.
GEN3i
I3763-3.1 en HBM: public
181