Environmental Specifications
Temperature Range
Operational
0 °C to +40 °C (+32 °F to +104 °F)
Non
-
operational (Storage)
-
25 °C to +70 °C (
-
13 °F to +158 °F)
Relative humidity
0% to 80%; non
-
condensing; operational
Protection class
IP20
Altitude
Maximum 2000 m (6562 ft) above sea level; operational
Shock: IEC 60068
-
2
-
27
Operational
Half
-
sine 10 g/11 ms; 3
-
axis, 1000 shocks in positive and negative direction
Non
-
operational
Half
-
sine 25 g/6 ms; 3
-
axis, 3 shocks in positive and negative direction
Vibration: IEC 60068
-
2
-
64
Operational
1 g RMS,
½
h; 3
-
axis, random 5 to 500 Hz
Non
-
operational
2 g RMS, 1 h; 3
-
axis, random 5 to 500 Hz
Operational Environmental Tests
Cold test IEC60068
-
2
-
1 Test Ad
-
5 °C (+23 °F) for 2 hours
Dry heat test IEC
-
60068
-
2
-
2 Test Bd
+40 °C (+104 °F) for 2 hours
Damp heat test IEC60068
-
2
-
3 Test Ca
+40 °C (+104 °F), humidity > 93% RH for 4 days
Non
-
Operational (Storage) Environmental Tests
Cold test IEC
-
60068
-
2
-
1 Test Ab
-
25 °C (
-
13 °F) for 72 hours
Dry heat test IEC
-
60068
-
2
-
2 Test Bb
+70 °C (+158 °F) humidity < 50% RH for 96 hours
Change of temperature test
IEC60068
-
2
-
14 Test Na
-
25 °C to +70 °C (
-
13 °F to +158 °F)
5 cycles, rate 2 to 3 minutes, dwell time 3 hours
Damp heat cyclic test
IEC60068
-
2
-
30 Test Db variant 1
+25 °C/+40 °C (+77 °F/+104 °F), humidity >95/90% RH
6 Cycles, cycle duration 24 hours
Harmonized Standards for CE Compliance, According to the Following Directives
Low Voltage Directive (LVD): 2014/35/EU
ElectroMagnetic Compatibility Directive (EMC): 2014/30/EU
Electrical Safety
EN 61010
-
1 (2010)
Safety requirements for electrical equipment for measurement, control, and laboratory use
-
General requirements
EN 61010
-
2
-
030 (2010)
Particular requirements for testing and measuring circuits
Electromagnetic Compatibility
EN 61326
-
1 (2013)
Electrical equipment for measurement, control and laboratory use
-
EMC requirements
-
Part 1: General requirements
Emission
EN 55011
Industrial, scientific and medical equipment
-
Radio
-
frequency disturbance characteristics
-
Limits and methods of
measurement
Radiated disturbance: class A
Immunity
EN 61000
-
4
-
2
Electrostatic discharge immunity test (ESD);
contact discharge
±
4 kV/air discharge
±
8 kV: performance criteria B
EN 61000
-
4
-
3
Radiated, radio
-
frequency, electromagnetic field immunity test;
80 MHz to 2.7 GHz using 10 V/m, 1000 Hz AM: performance criteria A
EN 61000
-
4
-
4
Electrical fast transient/burst immunity test
All input and output connections
±
2 kV using capacitive clamp: performance criteria B
EN 61000
-
4
-
6
Immunity to conducted disturbances, induced by radio
-
frequency fields
150 kHz to 80 MHz, 1000 Hz AM; 10 V RMS @ all inputs, both using clamp: performance criteria A
GEN3i
784
I3763-3.1 en HBM: public