background image

20

M408, M416

HBM

A2400-1.1 en/de

6

Specifications

Type

M408

M416

Number of input channels

4

Number of output channels

8

16

Actual spectrum change

1)

2 times, through opto−

switch

4 times, through opto−

switch

Relative sampling rate

2)

0.5 times

0.25 times

Insertion loss/channel

3)

dB

typ. < 3.0

typ. < 3.0

Dynamic range

dB

40

30

Sampling rate

1/s

max. 500

max. 250

Optical connector socket

FC/APC

Ambient conditions

Nominal temperature range
Ambient humidity range
Storage temperature range
Storage humidity range

°

C [

°

F]

% rel. h.

°

C [

°

F]

% rel. h.

0...+50 [+32...+122]

0...80, non-condensing

−20...+70 [−4...+158]

0...95, non-condensing

Electrical connection values

Supply voltage

User-defined interface to the interrogator through

8-pole DIN circular

Weight, approx.

kg

1.4

1)

Optical switches allow an actual increase of the measurement spectrum available by factor 2 to 4.

2)

The optical switches require that the multiplexer acquires data at alternating switch positions. This reduces
the interrogator’s relative sampling rate by a factor equalling the degree of multiplexing. The sampling rate
is maximally reduced by a factor of 2 when using the 8-channel expansion and maximally by a factor of 4
when using the 16-channel expansion module.

3)

Insertion loss data apply only for one direction (monomode).

Summary of Contents for M408

Page 1: ...A2400 1 1 en de Multiplexer for optical strain gages Multiplexer f r optische Dehnungsmessstreifen M408 M416 Operating Manual Bedienungsanleitung...

Page 2: ......

Page 3: ...n 12 2 3 Maintenance and cleaning 12 3 Connection 14 3 1 Multiplexer connection to the interrogator 14 3 2 Typical optical performance and qualification 16 3 3 Remote control interface 16 3 4 Connecti...

Page 4: ...erson instructed to carry out installation commissioning maintenance or repair of the device must have read and understood the Operating Manual and in particular the technical safety instructions Rema...

Page 5: ...pplication and disposal instructions as well as useful information Symbol NOTE Means that important information about the product or its handling is being given Symbol Meaning CE mark The CE mark enab...

Page 6: ...any repair or soldering work on motherboards is prohibited When exchanging complete modules use only original parts from HBM Qualified personnel This device is only to be installed and used by qualif...

Page 7: ...ver use optical instruments to look at it Even the reflection of laser light from a smooth surface can be harmful Do not direct the laser beam onto flammable gas or flammable liquids as this could cau...

Page 8: ...failure to comply with safety requirements can result in damage to property or destruction of the device The user must have read and understood all instructions before continuing to use Laser safety...

Page 9: ...o the optical connecting sockets For the location of the optical outputs see Chapter 2 The laser is activated when the device is connected to the mains and initialized The red LED on the front of the...

Page 10: ...t thermal and service life stability Seamless integration with associated interrogator 1 2 Applications Strain measurements in civil engineering including bridges roads and dams Continuous monitoring...

Page 11: ...n LED display is located on the bottom right of the front panel of the M4 device This LED shows the status of the multiplexer Red LED Ready for operation Lights up when the interrogator has been power...

Page 12: ...be operated in a pollution degree rating 2 area that is where normally only non conductive pollution occurs with only occasional temporary conductivity caused by condensation 2 3 Maintenance and clea...

Page 13: ...wards the edge of the compress End this movement in a clean and dry area of the compress Repeat the process making sure that you never use any parts of the compress that are dirty 5 Dispose of the dir...

Page 14: ...l inputs of the multiplexer must be connected to the applicable detector channel on the interrogator In other words the physical detector channel 1 of the interrogator should be connected with an APC...

Page 15: ...15 M408 M416 HBM A2400 1 1 en de Fig 3 1 Connection diagram of a multiplexer with a 4 channel DI interrogator...

Page 16: ...sheets of the applicable interrogators in the series SI and DI 3 4 Connecting the sensor chains The M408 and M416 devices are equipped with optical connector sockets for 8 or 16 optical sensor chains...

Page 17: ...he operating ambient temperature in the rack may be higher than the room temperature max 40 C 104 F Reduced air flow The device should be mounted in the rack in such a way that the air flow necessary...

Page 18: ...ttings must be made accordingly in catman Easy AP as described for the interrogators The increased number of channels due to the multiplexer must be set separately in catman Easy AP Connect the interr...

Page 19: ...gator 1 5 1 1 2 6 2 2 3 7 3 3 4 8 4 4 16 multiplex channels Group Interrogator 1 5 9 13 1 1 2 6 10 14 2 2 3 7 11 15 3 3 4 8 12 16 4 4 Table Allocation of multiplexer sensor chains One or more signals...

Page 20: ...32 122 0 80 non condensing 20 70 4 158 0 95 non condensing Electrical connection values Supply voltage User defined interface to the interrogator through 8 pole DIN circular Weight approx kg 1 4 1 Op...

Page 21: ...21 M408 M416 HBM A2400 1 1 en de 7 Dimensions in mm 1 mm 0 03937 inches 234 123 114...

Page 22: ...22 M408 M416 HBM A2400 1 1 en de...

Page 23: ...3 Wartung und Reinigung 32 3 Anschlie en 34 3 1 Multiplexeranschluss an den Interrogator 34 3 2 Typische optische Leistung und Qualifikation 36 3 3 Fernsteuerungsschnittstelle 36 3 4 Sensorketten ansc...

Page 24: ...em Per sonal unsachgem eingesetzt und bedient wird Jede Person die mit Aufstellung Inbetriebnahme Wartung oder Reparatur des Ger tes beauftragt ist muss die Bedienungsanleitung und insbesondere die si...

Page 25: ...te Symbole f r Anwendungs und Entsorgungshinweise sowie n tzliche Informa tionen Symbol HINWEIS Weist darauf hin dass wichtige Informationen ber das Produkt oder ber die Handhabung des Produktes gegeb...

Page 26: ...ter sagt Bei Austausch gesamter Baugruppen sind nur Originalteile von HBM zu verwenden Qualifiziertes Personal Dieses Ger t ist nur von qualifiziertem Personal ausschlie lich entsprechend der technisc...

Page 27: ...strahl und betrachten Sie diesen niemals mit optischen Instrumen ten Auch die Reflektion des Laserlichts von einer glatten Oberfl che aus kann sch dlich sein Richten Sie den Laserstrahl nicht auf entz...

Page 28: ...e haben kann Vor dem weiteren Einsatz muss der Nutzer alle Hinweise gelesen und verstanden ha ben Laser Sicherheit Information zur Sicherheit des durchstimmbaren Laser Frequenzgenerators Ger tespezifi...

Page 29: ...er Laser ist aktiviert wenn das Ger t an das Netz angeschlossen und initiali siert ist Die rote LED auf der Ger tevorderseite zeigt an ob die Netzversor gung eingeschaltet ist Schauen Sie keinesfalls...

Page 30: ...chnete Temperatur und Langzeitbest ndigkeit D Nahtlose Integration mit zugeh rigem Interrogator 1 2 Anwendungen D Dehnungsmessungen im Tiefbau einschlie lich Br cken Stra en und Stau d mme D Kontinuie...

Page 31: ...echten Seite der Frontplatte der Ger te M4 ist eine LED Anzeige angeordnet Diese LED zeigt den Status des Multiplexers an Rote LED Funktionsbereit Leuchtet nachdem der Interrogator hochgefahren ist un...

Page 32: ...h Auftreten von blicherweise nur nicht leitf higer Verschmutzung bei gelegentlicher vor bergehender Leitf higkeit durch Betauung 2 3 Wartung und Reinigung Der Multiplexer ist wartungsfrei Beachten Sie...

Page 33: ...und benutzen Sie auf keinen Fall verschmutzte Teile der Kompresse 5 Entsorgen Sie die verschmutzte Kompresse HINWEIS Die optischen Anschlussbuchsen CH1 CH16 f r die Sensorketten m ssen unbedingt saube...

Page 34: ...optischen Eing nge des Multiplexers sollte an den entsprechen den Detektorkanal am Interrogator angeschlossen werden Anders ausge dr ckt sollte der physische Detektorkanal 1 des Interrogators mit eine...

Page 35: ...35 M408 M416 HBM A2400 1 1 en de Abb 3 1 Anschlussschema eines Multiplexers mit einem 4 Kanal DI Interrogator...

Page 36: ...a tenbl tter des entsprechenden Interrogators der Baureihen SI und DI 3 4 Sensorketten anschlie en Die Ger te M408 bzw M416 sind mit optischen Anschlussbuchsen f r 8 bzw 16 optische Sensorketten ausge...

Page 37: ...e Rack kann die Betriebs Umgebungstemperatur im Rack h her sein als die Raumtemperatur max 40 C Verringerter Luftstrom Die Montage des Ger ts im Rack sollte so erfol gen dass der f r den sicheren Betr...

Page 38: ...die Einstellungen in catman Easy AP wie f r die Interroga toren beschrieben vorgenommen Die erweiterte Kanalzahl durch die Multiple xer muss in catman Easy AP separat eingestellt werden Verbinden Sie...

Page 39: ...6 2 2 3 7 3 3 4 8 4 4 16 Multiplexkan le Gruppe Interrogator 1 5 9 13 1 1 2 6 10 14 2 2 3 7 11 15 3 3 4 8 12 16 4 4 Tabelle Zuordung der Multiplexer Sensorketten Je nach St rke des Signals und Einste...

Page 40: ...nicht kondensierend Elektrische Anschlusswerte Versorgungsspannung Benutzerdefinierte Schnittstelle zum Interrogator ber 8 pol DIN Rundstecker Gewicht ca kg 1 4 1 Optische Schalter erm glichen eine e...

Page 41: ...41 M408 M416 HBM A2400 1 1 en de 7 Abmessungen 234 123 114...

Page 42: ...42 M408 M416 HBM A2400 1 1 en de...

Page 43: ...43 M408 M416 HBM A2400 1 1 en de...

Page 44: ...1 1 en de nderungen vorbehalten Alle Angaben beschreiben unsere Produkte in allgemeiner Form Sie stellen keine Beschaffenheits oder Haltbarkeitsgarantie im Sinne des 443 BGB dar und begr nden keine Ha...

Reviews: