17
ME 30
4.6
Measuring Frequency Range / Cut-off Frequency f
co
The Butterworth low pass of the amplifier has been set to 0...60Hz (–1dB) by
the manufacturer. It can be readjusted by switching the DIP switch S24/3 to
0...2Hz (–1dB).
Measuring frequency range
S 24/3
0...60Hz (–1dB); (ex-works setting)
f
co
= 80Hz (–3dB)
0...2Hz (–1dB);
f
co
= 2.5Hz (–3dB)
4.7
Synchronization
When operating several ME30 amplifiers (in a common housing), one of them
has to be switched as the master the carrier frequency of which supplies the
clock pulses, the others are to be switched as slave. Pins 11c of all amplifiers
have to be joined for synchronization. The master amplifier is selected by
switch S28.
Synchronization
Switch S28
Slave amplifier (ex-works setting)
Master amplifier
Summary of Contents for ME30