9
ME 30
2
Electrical Connection
2.1
Transducer Connection
Strain gauge transducers are connected by means of 6-wire circuits.
With transducers in 4-wire technology the sensing leads are not needed. In
this case the contacts 32a+c and 28a+c on the connector must be linked.
screen
pins 31a+c
. . . . . . . . . . . . . . . . . .
rd
gy
bk
wh
bu
gn
ye
measuring signal (–)
pin 30a
. . . . . .
sensing line to 32a (–)
pin 32c
. . . . .
bridge supply voltage (–)
pin 32a
. .
measuring signal (+)
pin 30c
. . . . . .
bridge supply voltage (+)
pin 28a
. .
sensing line to 28c (+)
pin 28c
. . . . .
contact
(alternatively 29a+c)
Fig. 2.1:
Bridge connection
The required bridge excitation voltage set by the factory is 5V(symmetrical
about ground). It can be switched to 2.5V by switch S25.
U
B
Bridge resistance
Switch S25
2.5V
R
B
≥
60...4000
Ω
5V
(ex-works settings)
R
B
≥
110...4000
Ω
The safety barriers SD01 must be wired into the connecting lines (see chap-
ter 5.4) for transducers in EEx(i) design.