Internal calculation channels
PMX
A4354-2.2
HBM: public
189
Outputs
Extreme Input 1
Input 2 at Extr.1
New extreme value
Parameters
-
Type
[Maximum, Minimum
,
Peak-to-Peak]
-
Inversion of
the hold
input
- Discharge rate [unit/second]
Default
Type: Maximum
Digital inputs: Not in use
Inversion: off
Discharge rate : 0.0
Exception handling
If the input is marked as invalid (Invalid Bit), then the output is
also marked as invalid.
The extreme value is still determined and output.
When the Input 2 is marked as invalid (invalid bit), the value is
captured nonetheless and also marked as invalid.
Example: Capturing displacement when F = F
max
:
Force F [N]
Displacement d [mm]
Monitored for
maximum values
the captured input
the new peak flag
This d-value is hold
until next peak
This d-value is hold
until next peak
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...