Communication with a control system
280
A4354-2.2
HBM: public
PMX
Storing data in the device
The file is stored in public device storage. It can be down
loaded or deleted there. Access via menu
System – Device – Device storage – Show device
storage)
Folder
public/PROFINET or
public/EtherCAT® or public/EtherNet/IP.
The folder can also be displayed in the web browser. To
do this enter the following in the address line:
http://pmx/public/PROFINET/ or
http://pmx/public/EtherCAT® or
http://pmx/public/EtherNet/IP
where "pmx" is the PMX network name. Names are case
sensitive!
Alternative notation with the IP address, for example
http://172.19.201.184/public/PROFINET/
Exceptions
The file is created from a template file residing in public/
PROFINET or public/EtherCAT® or public/EtherNet/IP. If
that file is not found, an error message appears "Cannot
open source file."
A firmware update (including one with the same version
number as the installed firmware) will restore the tem
plate file.
If a measurement card is not installed and no calculated
channels have been transferred yet, the generated file is
not a valid file and will not be accepted by PROFINET
configurators or. EtherCAT® or EtherNet/IP .
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...