PMX command set (API)
412
A4354-2.2
HBM: public
PMX
Response:
p1 : 1
q1
(
y)
q1: sum of binary representation of single statusbits of
Status 1, see table below
Status 1
binary Value
Remarks
No error
0
FACTORYSETTINGS_ERROR
1
factory calibration for PMX
housing
corrupted (not meascards !)
SYNC_MASTER
2
bit set:sync master bit cleared: sync
slave
SYNCMESSAGE_ERR
4
missing or invalid sync
messages
SYNC_UNLOCKEDSLAVE_ERR
8
cannot lock, regulator is not able to
synchronize to master
ALIVE
16
toggles with approx
.
1Hz
POWEROVERLOAD
32
power supply overload somewhere
CAT_BUF_OVERRUN
64
“Catman” interface buffer overrun
(TSV..) ==> meas. value stream
“broken”
SYSTEM_NOT_READY
128
e
.
g
.
change of parameterset in pro
gress or failed
DSP_OVERRUN
256
e
.
g
.
too many calculated channels
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...