PMX command set (API)
PMX
A4354-2.2
HBM: public
451
Calibration Dead Load Target
Zero point offset target value
Enter the target value for the zero point offset of the input
characteristic (for command CDV) for all selected
channels (PCS/SPS).
Syntax:
CDTp1(x)
Parameters:
p1:The target value should be set to the current
measured value
Effect:
The value in displayed units, to which the amplifier should
be set by the CDV command (no parameters). The
factory setting is 0.
Note:
The response depends on the SRB command.
Calibration Dead Load Target Query
Output the zero point offset target value
Output the target value for the zero point offset of the
input characteristic (for command CDV) for all selected
channels (PCS/SPS).
Syntax:
CDT?(x)
Parameters:
none
Response:
q1(y):
Target value, to which the current measured value
is set e. g. 0.01
,
0
,
0.5
,
0.502
Virtual subchannels (channel 9) do not support "dead
load targets". q1=0.
CDT
CDT?
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...