FAQs
PMX
A4354-2.2
HBM: public
535
Depending on how many sub-parameter sets are
switched, the switching time takes between 0.1 and
2.5 seconds.
S
Can the PMX also be used as a fieldbus master?
PMX can only be used as a slave in the Ethernet-
based fieldbuses (EtherCAT, PROFINET and
EtherNet/IP).
If the CANopen interface is used under CODESYS
Soft-PLC, the PMX can be run either as a CAN master
or CAN slave.
S
What happens if the supply voltage fails during the
saving of a parameter set?
This will destroy the parameter set and the PMX will
boot up with the factory setting after being switched
back on again. To avoid this, backing up the device
settings to PC is urgently recommended.
S
What happens if the supply voltage fails during a
firmware update?
The device either comes online again with the old
firmware after being switched on or the device loads
and installs the new/loaded firmware and is then
ready for operation after approx. 10-15 minutes.
S
What happens with a CODESYS application after a
power failure?
If the CODESYS project was saved as a "boot pro
ject," the application will automatically start running
again after the PMX is turned on.
S
Can the source code of a CODESYS project be re
loaded from the PMX?
No, because the compilation generates machine
code, which is loaded into the PMX. This ensures
knowhow protection.
However, when a program is created, the original
source code can also be transferred to the PMX and
Summary of Contents for PX401
Page 1: ...Operating Manual English PMX...
Page 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Page 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Page 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Page 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Page 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Page 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Page 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Page 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Page 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Page 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Page 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Page 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Page 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Page 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Page 489: ...Object directory PMX A4354 2 2 HBM public 489...
Page 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...